Skip to main content
SHARE
Publication

A Multi-Anode Photomultiplier Tube Based Wavelength-Shifting-Fiber Detector for neutron diffraction...

Publication Type
Conference Paper
Publication Date
Page Numbers
180 to 184
Conference Name
2012 IEEE Nuclear Science Symposium
Conference Location
Anaheim, California, United States of America
Conference Date
-

The wavelength-shifting (WLS) fiber scintillator neutron detectors were developed for two time-of-flight (TOF) neutron powder diffractometers (POWGEN, VULCAN) at Spallation Neutron Source (SNS). In a recent module (v3.0), however, there are 32 1-inch-diameter photomultiplier tubes (PMTs) which are bulky and expensive. We built a new detector module (v3.1) based on four multi-anode (MA) PMTs, and tested its performance including detection efficiency, count rate capability, spatial resolution, ghosting properties, and gamma-ray sensitivity. The v3.1 module was compared with two prior v3.0 modules, and 3He tube detectors.