Skip to main content
SHARE
Publication

Combining Transmission Geometry Laser Ablation and a Non Contact Continuous Flow Surface Sampling Probe/Electrospray Emitter ...

by Olga S Ovchinnikova, Vilmos Kertesz, Gary J Van Berkel
Publication Type
Journal
Journal Name
Rapid Communications in Mass Spectrometry
Publication Date
Page Numbers
3735 to 3740
Volume
25
Issue
24

This paper describes the coupling of ambient pressure transmission geometry laser ablation with a liquid phase sample collection into a continuous flow surface sampling probe/electrospray emitter for mass spectrometry based chemical imaging. The flow probe/emitter device was placed in close proximity to the surface to collect the sample plume produced by laser ablation. The sample collected was immediately aspirated into the probe and on to the electrospray emitter, ionized and detected with the mass spectrometer. Freehand drawn ink lines and letters and an inked fingerprint on microscope slides were analyzed. The circular laser ablation area was about 210 m in diameter and under the conditions used in these experiments the spatial resolution, as determined by the size of the surface features distinguished in the chemical images, was about 100 μm.