February 2012

Journal

Position Sensitive Detection System for Charged Particles

By:
Coello, E.; Favela, F.; Curiel, Q.; Chavez, E; Huerta, A.; Varela, A.; Shapira, Dan
Journal Name:
Revista Mexicana de Fisica
Page Number:
198-204
Volume:
58
Issue Number:
3
Publication Date:
February 2012

Abstract

The position sensitive detection system presented in this work employs the Anger logic algorithm to determine the position of the light spark produced by the passage of charged particles on a 170 x 170 x 10 mm3 scintillator material (PILOT-U). The detection system consists of a matrix of nine photomultipliers, covering a fraction of the back area of the scintillators. Tests made with a non-collimated alpha particle source together with a Monte Carlo simulation that reproduces the data, suggest an intrinsic position resolution of up to 6 mm is achieved.