The increase in neutron flux at the Oak Ridge National Laboratory (ORNL) Spallation Neutron Source (SNS), currently operating at 2.0 MW proton beam power, has created new opportunities for higher-throughput neutron scattering experiments while also increasing the importance of minimizing background scattering and optimizing sample-environment operations. To address these challenges on the Backscattering Silicon Spectrometer (BA-SIS), several upgrades were developed and evaluated, including boron carbide (B₄C) masking for flat-plate sample containers, multi-cell sample holders used with a vertically translating sample stick, and an automated helium pump and purge (HPP) system for closed-cycle refrigerators. Neutron diffraction measurements demonstrate that B₄C masks reduce background scattering by 49–67%, outperforming both borated aluminum and boron nitride masks while introducing no additional Bragg reflections within the instrument’s accessible Q-range. Commissioning tests of a double-cell flat-plate sample contain-er showed no measurable crosstalk between adjacent sample compartments and con-firmed stable thermal performance, enabling multiple samples to be measured without re-peated temperature cycling. In addition, the automated HPP system provided reproducible sample-space gas handling with approximately ±1 mbar precision while reducing the need for operator intervention and supporting remote operation. Together, these developments improve signal-to-noise performance, increase experimental throughput, and enhance operational efficiency at BASIS, supporting the instrument’s continued operation under higher neutron flux conditions.