Correcting aberrations of a transverse-field neutron resonance spin echo instrument

By:
Kuhn, Stephen J; McKay, Samuel R; Li, Fankang ; Dalgliesh, Robert; Dees, Eric B; Burrage, Kaleb C; Shen, Jiazhou; Pynn, Roger
Journal Name:
Review of Scientific Instruments
Page Number:
033901-1-033901-8
Volume:
94
Issue Number:
3
Publication Date:
June 2025
View DOI Listing:
https://doi.org/10.1063/5.0128097

Abstract

Neutron resonance spin echo (NRSE) technique has the potential to increase the Fourier time and energy resolution in neutron scattering by using radio frequency (rf) neutron spin-flippers. However, aberrations arising from variations in the neutron path length between the rf flippers reduce the polarization. Here, we develop and test a transverse static-field magnet, a series of which are placed between the rf flippers, to correct for these aberrations. The prototype correction magnet was both simulated in an NRSE beamline using McStas, a Monte Carlo neutron ray-tracing software package, and measured using neutrons. The results from the prototype demonstrate that this static-field design corrects for transverse-field NRSE aberrations.