August 2009

Conference Paper

Development of a Multispectral Tissue Characterization System for Optimization of an Implantable Perfusion Status Monitor for Transplanted Liver

By:
Baba, Justin S; Letzen, Brian S; Ericson, Milton N; Cote, Gerard; Xu, Weijian; Wilson, Mark
Journal Name:
Engineering in Medicine and Biology Society Annual Conference
Page Number:
6565-6568
Volume:
2009
Issue Number:
N/A
Publication Date:
August 2009
Conference Name:
31st Annual International Conference of the IEEE EMBS
Conference Location:
Minneapolis, Minnesota, United States of America
View DOI Listing:
https://doi.org/10.1109/IEMBS.2009.5334499

Abstract

Optimizing wavelength selection for monitoring perfusion during liver transplant requires an in-depth characterization of liver optical properties. With these, the impact of liver absorption and scattering properties can be investigated to select optimal wavelengths for perfusion monitoring. To accomplish this, we are developing a single integrating-sphere-based using a unique spatially resolved diffuse reflectance system for optical properties determination for thick samples. We report early results using a monochromatic source implementation to measure the optical properties of well characterized tissue phantoms made from polystyrene spheres and Trypan blue. The presented results show the promise of using this unique system to measure the optical properties of the tissue phantoms. We are currently in the process of implementing an automated Levenberg–Marquardt fitting algorithm to determine the peak location of the diffuse reflectance profile to ensure robust computation of sample optical properties. Future work will focus on the incorporation of multispectral capability to the technique to facilitate development of more realistic liver tissue phantoms.


Related Researchers