Particle Loss in a Scanning Mobility Particle Analyzer Sampling Extension Tube

By:
Hollenbeck, Scott M; Jankovic, John ; Ogle, Burton R; Zontek, Tracy ; Hall, Michaela A
Journal Name:
International Journal of Occupational and Environmental Health
Page Number:
429-433
Volume:
16
Issue Number:
4
Publication Date:
December 2010

Abstract

Deposition of particles in sampling lines may occur due to inertial, electrostatic, or diffusional forces among others. Very small particles such as those in the nanoscale1, are not susceptible to losses from inertial deposition.2 Electrical effects are reportedly controlled by the use of conductive tubing.2 Particle losses from Brownian diffusion affect both particle size distribution and number concentration. Selectively removing the smallest particles has the effect of increasing the statistical measures of particle size; mean, median, geometric mean, mode, while decreasing number concentration and geometric standard deviation. These losses if quantified can be used to interpret or correct the data.3,4 Sample loss from either a rigid graphitic or flexible Tygonâ„¢ tube attached to a Scanning Mobility Particle Sizerâ„¢ inlet was investigated as part of the routine sampling process at the Center for Nanophase Materials Sciences. Mean concentrations and particle size parameters determined from samples collected with and without sample inlet extensions of both types were compared. No losses or changes in aerosol size distribution parameters that required correction were found for either material at lengths of ~0.7m.