Skip to main content
SHARE
Publication

Vacuum System Upgrade for Extended Q-Range Small-Angle Neutron Scattering Diffractometer (EQ-SANS) at SNS...

by Christopher M Stone, Derrick C Williams, Jeremy P Price
Publication Type
Journal
Journal Name
MethodsX
Publication Date
Page Numbers
525 to 534
Volume
3

The Extended Q-Range Small-Angle Neutron Scattering Diffractometer (EQ-SANS) instrument at the Spallation Neutron Source (SNS), Oak Ridge, Tennessee, incorporates a 69m3 detector vessel with a vacuum system which required an upgrade with respect to performance, ease of operation, and maintenance. The upgrade focused on improving pumping performance as well as optimizing system design to minimize opportunity for operational error. This upgrade provided the following practical contributions:
• Reduced time required to evacuate from atmospheric pressure to 2mTorr from 500-1,000 minutes to 60-70 minutes
• Provided turn-key automated control with a multi-faceted interlock for personnel and machine safety.