October 2012

Conference Paper

Wavelength-Shifting-Fiber Scintillation Detectors for Thermal Neutron Imaging at SNS

By:
Clonts, Lloyd G; Cooper, Ronald G; Crow Jr, Morris L; Diawara, Yacouba ; Ellis, Edward D; Funk, Loren L; Hannan, Bruce W; Hodges, Jason P; Richards, John D; Riedel, Richard A; Wang, Cailin
Page Number:
4877-4882
Publication Date:
October 2012
Conference Name:
IEEE Nuclear science Symposium and Medical Imaging Conference-2011
Conference Location:
Valencia, Spain

Abstract

We have developed wavelength-Shifting-fiber Scintillator Detector (SSD) with 0.3 m2 area per module. Each module has 154 x 7 pixels and a 5 mm x 50 mm pixel size. Our goal is to design a large area neutron detector offering higher detection efficiency and higher count-rate capability for Time-Of-Flight (TOF) neutron diffraction in Spallation Neutron Source (SNS). A ZnS/6LiF scintillator combined with a novel fiber encoding scheme was used to record the neutron events. A channel read-out-card (CROC) based digital-signal processing electronics and position-determination algorithm was applied for neutron imaging. Neutron-gamma discrimination was carried out using pulse-shape discrimination (PSD). A sandwich flat-scintillator detector can have detection efficiency close to He-3 tubes (about 10 atm). A single layer flat-scintillator detector has count rate capability of 6,500 cps/cm2, which is acceptable for powder diffractometers at SNS.