August 2021

Journal

Monte Carlo Analysis of the Performance of the ITER Diagnostic Residual Gas Analyzer

By:
Klepper, Conrad ; Ravelli, Fabio A
Journal Name:
Fusion Science and Technology
Page Number:
1-12
Volume:
00
Issue Number:
XXXX 2021
Publication Date:
August 2021
View DOI Listing:
https://doi.org/10.1080/15361055.2021.1898867

Abstract

The composition of exhausted gas is a key parameter in long-pulse plasma fusion experiments, and its evolution shall be monitored at timescales relevant to plasma dynamics and plasma-wall interactions. A diagnostic residual gas analyzer (DRGA) is a multisensor instrument particularly suited to these studies, and ITER will adopt DRGAs in the equatorial and in the divertor tokamak regions. In this work, we have revisited the design of the ITER divertor DRGA through simple vacuum analytical considerations supported by simulations conducted with Molflow+, a test particle Monte Carlo (TPMC) simulation code commonly used in the particle accelerator community. Starting with recommendations on the manufacturing of the vacuum piping of the DRGA, this work is followed by a complete vacuum characterization of the diagnostic vacuum setup (pressure profiles at base pressure and during sampling, orifice diameter, and length optimization), and finally, the in-vessel residence time of the most important gas species is simulated. These studies have allowed us to give insights into some experimental results recently found on the prototype DRGA installed in the Wendelstein W7-X stellarator.