- By:
- Korsah, Kofi ; Murray, William ; Tomkins, Bruce A; Kertesz, Vilmos
- Publication Date:
- September 2008
Abstract
Oak Ridge National Laboratory (ORNL) was contracted by the United States Air Force Metrology Calibration Unit (AFMETCAL) to investigate the effect of relative humidity and carbon dioxide on the Interscan model 4187/4189 hydrazine monitors in the U.S. Air Force inventory. The evaluations were performed with a KinTek 491 reference gas generation system, which provided humidified nitrogen [or air with certified ppm concentrations of carbon dioxide (CO2)] containing various concentrations of hydrazine. In several of the tests, both the 4187 and the 4189 were used interchangeably because both use the same type of electrochemical sensor. The major difference is in the scale readings. The model 4187 has two ranges: one scale has a range of 0 to 100 ppb, at 2ppb/division. The other scale has a range of 0 to 2000ppb at 40ppb/division. The model 4189 also has two ranges: one scale has a range of reads 0 to 100 ppb, at 2ppb/division. The other scale has a range of 0 to 1000ppb at 20ppb/division. For the relative humidity (RH) tests, the RH of each of three different reference gas hydrazine concentrations was varied over the range of 5 to 65%. At each relative humidity value, the Interscan monitor was used to read the hydrazine concentration; while coulometric titration was used to determine the hydrazine concentration of a sample of the gas. These tests showed that the response of the Interscan 4187 is not affected by changes in humidity in the range of 5 to 65%. With respect to the CO2 tests, preliminary test results indicated poor performance (low readings in high CO2 environments) that could be attributable to very long response times. Three tests were performed to verify this. In the first test, the KinTek 491 reference gas generation system was used to generate a reference gas containing 80 ppb of hydrazine and 720 ppm of carbon dioxide in air. The response of the sensor in the Interscan 4189 was then measured with time. In the second test, the KinTek 491 reference gas generation system was used to generate a reference gas containing 80 ppb of hydrazine and 476 ppm of carbon dioxide in air. The response of the sensor in the Interscan 4189 was again measured with time. The third and final test was similar to the first two as described above, except that the reference gas was hydrazine mixed with pure nitrogen containing no C02. The results showed that the Interscan monitor�s response time was adversely affected by the CO2 concentration. In particular, the response time increased with increase in the concentration of the CO2. For a reference gas containing 80 ppb of hydrazine and 720 ppm of CO2 in air, it took 90 minutes for the monitor�s reading to reach 40 ppb (50% response as used by Interscan for calibrations); and it took 115 minutes for the reading to reach 65 ppb (which is within the 80 ppb + 15 ppb specifications of the sensor). For the reference gas containing 80 ppb of hydrazine and 476 ppm of CO2 in air, it took only 18 minutes for the monitor�s reading to reach 40 ppb (50% response) and 40 minutes for it to reach 65 ppb. For the tests using hydrazine mixed with pure nitrogen, the measured response reached 40 ppb (50% response) in 1 minute or less for all three runs. In addition, all three runs reached at least 72 ppb (90 % response) within 10 minutes. Furthermore, the reading was 80 ppb + 15 ppb and holding at the same reading (100% response) from 12 minutes to 20 minutes for all three runs. All three runs showed the readings return to zero within a minute or less of being disconnected from the 84 ppb hydrazine reference gas. These response times are faster than Interscan�s specified response time for the Interscan 4189.