- By:
- Van Berkel, Gary J; Kertesz, Vilmos
- Journal Name:
- Rapid Communications in Mass Spectrometry
- Page Number:
- 1380-1386
- Volume:
- 23
- Issue Number:
- 9
- Publication Date:
- May 2009
- View DOI Listing:
- https://doi.org/10.1002/rcm.4014
Abstract
This paper reports on the conversion of a liquid microjunction surface sampling probe (LMJ-SSP) into a two electrode electrochemical cell using a conductive sample surface and the probe as the two electrodes with an appropriate battery powered circuit. With this LMJ-SSP, two-electrode cell arrangement, tagging of analyte thiol functionalities (in this case peptide cysteine residues) with hydroquinone tags was initiated electrochemically using a hydroquinone doped solution when the analyte either was initially in solution or was sampled from a surface. Efficient tagging (~90%), at flow rates of 5-10 �L/min, could be achieved for up to at least two cysteines on a peptide. The high tagging efficiency observed was explained with a simple kinetic model. In general, the incorporation of a two-electrode electrochemical cell, or other multiple electrode arrangement, into the LMJ-SSP is expected to add to the versatility of this approach for surface sampling and ionization coupled with mass spectrometric detection.