Skip to main content
SHARE
Publication

Experimental and computational characterization of a modified Sioutas Cascade impactor for respirable radioactive aerosols

by Yadukrishnan Sasikumar, Vineet Kumar, Rosemary A Montgomery, Prashant K Jain
Publication Type
Journal
Journal Name
Atmosphere
Publication Date
Page Numbers
156 to 180
Volume
16
Issue
2

Oak Ridge National Laboratory is collecting and characterizing aerosols released when spent nuclear fuel (SNF) rods are fractured in bending. An aerosol collection system was designed and tested to collect respirable sized (<10 μm aerodynamic diameter [AED]) particulates inside a hot cell facility. The setup is a modified version of the commercially available Sioutas cascade impactor, to which additional stages were added to expand the aerosol collection range from 2.5 to ~15 μm AED. To accommodate the additional stages and specific test conditions, the operating flow rate for aerosol collection was reduced, and testing was conducted by using pressure drop measurements, surrogate dust collection, and particle size characterization. The fluid flow distribution within the cascade and its stages was simulated in STAR-CCM+, and the stage-wise pressure drops obtained using the computational fluid dynamics model were then compared to experimental data. Lagrangian particle simulations were also performed, and stage-wise collection statistics were obtained from the simulation for comparison with the experimental data obtained using SNF-surrogate dust particles. The results provide valuable insights into the stage-wise particle collection characteristics of the modified cascade impactor and can also be used to improve the prediction accuracy of the manufacturer-determined analytical correlations.