Pressure/Vacuum

Bob Quinn completes an air velocity measurement        Bob Quinn performs a mass flow calibration

ORNL Metrology Laboratory (ORNLML) maintains a variety of pressure and vacuum standards traceable to SI (International System of Units) through the National Institute of Standards and Technology (NIST) and other international primary laboratories. These standards generate and measure various pressure and vacuum properties required to perform traceable calibrations of our customers’ instruments and equipment.  ORNLML can provide accredited calibrations for any pressure parameters within the ranges detailed in our Scope of Accreditation. However, traceable non-accredited calibrations are available for pressure and vacuum outside of our Scope of Accreditation. The tables below list our accredited capabilities. Our Scope of Accreditation includes complete details of our accredited capabilities.

Typical types of equipment the ORNLML performs pressure and vacuum calibrations on are: Pressure gauges and transducers, barometers, manometers, capacitance gauges and vacuum gauges.

If you have questions about our pressure and vacuum capabilities, please contact the ORNL Metrology Laboratory for more information.

Capability
Pressure –Absolute Mode (20/T05)

Range

Best Uncertainty (+/-) in ppm of Reading + mPa (psi)*

Remarks

1.0 Pa to 15.0 kPa
0.0000145 psia to 2.17 psia

37 + 8
37 + (1.2 E-6)

Source/Measure

 

Range

Best Uncertainty (+/-) in ppm of Reading + Pa*

Remarks

10 to 380 kPa
(1.45 to 55 psia)

15.7 + 0.10

Source/Measure

 

 

 

50 to 1900 kPa
(7.3 to 275 psia)

18.0 + 0.12

Source/Measure

 

 

 

200 to 7600 kPa
(29 to 1100 psia)

19.9 + 0.25

Source/Measure

Absolute Differential Mode

Range

Best Uncertainty (+/-) in ppm of Reading + mPa (psi) *

Remarks

0 Pa to 15 kPa
0 psia to 2.17 psia

37 + 5
37 + (7.3 E-7)

Source/Measure

Gauge Mode

Range

Best Uncertainty (+/-) in ppm of Reading + mPa (psi) *

Remarks

0 Pa to 15.0 kPa
0 psi to 2.17 psi

37 + 5
37 + (7.3 E-7)

Source/Measure

 

Range

Best Uncertainty (+/-) in ppm of Reading *

Remarks

10 to 380 kPa
(1.45 to 55 psi)

16.0 + 0.01 Pa

Source/Measure

 

 

 

50 to 1900 kPa
(7.3 to 275 psi)

18.3 + 0.06 Pa

Source/Measure

 

 

 

200 to 7600 kPa
(29 to 1100 psi)

20.2 + 0.23 Pa

Source/Measure

 

 

 

1.379 MPa to 68.948 MPA
(200 psig to 10 000 psig)

108

Source/Measure

*Represents an expanded uncertainty using a coverage factor, k=2, at an approximate level of confidence of 95%.

 